Astron. Astrophys. 359, 682-694 (2000)
6. Results
In this Sect. the methods explained in Sect. 5 are applied to the data
set discussed in Sect. 4.
6.1. and spectra
Fig. 5 shows the and
spectra. These spectra display a
relatively sharp knee at values consistent with a primary energy for
the knee as determined below.
6.2. Energy spectra
Fig. 6 displays the integral energy spectra uncorrected for an A
dependent bias obtained with the four reconstruction methods. The
differences in absolute normalisation and spectral slope originate
from the different mass dependent biases. After the correction of the
chemical bias the integral spectra are similar (Fig. 7). This
nontrivial fact is in favour of the internal consistency of data
analysed here; a longitudinal shower development different from the
one predicted by the Monte Carlo or errors in the calibration of
and
could have spoiled the agreement of
spectra obtained with different energy reconstructions. The
differential energy spectrum is shown in Fig. 8. A steepening of the
energy spectrum is visible around an energy of 4 PeV. There seems to
be no "fine structure" in the energy spectrum around the knee in
excess of 20 . Apparent structure with
smaller amplitudes that appears in the spectrum reconstructed with a
given energy-reconstruction method is not reproduced with other
methods. This is expected due to the A dependent bias of our
energy-reconstruction methods (see Fig. 2). Note that with these
methods, a potential structure in the energy spectrum consisting of
different nuclei is smeared out. If two different power laws, smoothly
connected at the knee (corresponding to a "sharp" knee), are fitted to
the differential spectra we obtain:
The reduced values of the fits to
the differential spectrum (12 d.o.f.) were 6.75, 4.03, 3.53 and 1.47
with energy reconstruction methods 1 to 4 respectively. Some of these
values are much larger than one. It is then difficult to specify a
statistical error; we specify the statistical errors for method 4 that
has a marginally acceptable reduced
value. The large values for the
analysis with energy-reconstruction methods 1-3 can be interpreted as
an argument in favour of a knee not absolutely sharp in energy.
However, the fact that one of the fits is acceptable on the 90
confidence level means that we cannot
reject the hypothesis of a "sharp" knee (two power laws with no
transition region) within our systematic errors. The large statistical
error on the knee position further indicates that we cannot reject the
hypothesis of a spectrum without a knee in the limited energy range of
this analysis with high significance.
The spectral index for the spectrum below the knee is consistent
with direct measurements at lower energies (Wiebel-Sooth et al. 1998)
and a recent Cherenkov-light based determination of the spectral index
in the TeV range (Aharonian et al. 1999); there is therefore no
evidence for any change in spectral index from the TeV range right up
to the knee.
6.3. Composition of CR
The fraction of light nuclei as a function of reconstructed energy -
obtained from the fits to the measured penetration-depth distributions
(see Sect. 5.2)- is presented in Fig. 10. At energies below the knee
the composition is mixed and consistent with direct measurements
around 100 TeV, namely ( )/all = 0.54
0.08 (Watson 1997). The data points
seem to indicate a gradual enrichment of heavy elements above about 1
PeV though the error bars are large (remember that there are only six
independent data points). We will discuss in Sect. 7 how
reliable the qualitative conclusion of a gradual enrichment in heavy
elements is within our systematic errors. The data rule out a
predominantly light composition at all energies and does not give
evidence for a drastic change of composition at the knee.
6.4. Elongation rate
Fig. 11 shows the corrected mean shower maximum depth as a function of
energy. A least-squares fit to the
values as a function of energy, using only the statistical errors,
![[EQUATION]](img97.gif)
yields an elongation rate ER=78.3
1.0 (stat) 6.2 (syst)
g/cm2 and mean depth parameter ERB=243.1
2.6 (stat)
15.7 (syst) g/cm2. The
specified mean values and statistical errors are the mean of fit
values with the four energy-reconstruction methods. The systematic
error is estimated as the standard deviation of the mean values
inferred with the four energy-reconstruction methods. The systematic
error introduced by the systematic uncertainty in slope is
smaller (about 3 and 14 g/cm2 for ER and ERB respectively).
The reduced values of the fit to
relation (6) (4 d.o.f.) are very large (6.6,9.2,17.2,23.5) for
energy-reconstruction methods 1-4, i.e the systematic errors dominate
over the rather small statistical errors for the mean
. Therefore the specified estimates
of the statistical errors obtained with the procedure explained in
Sect. 5.4 have to be treated with caution. The data point at the
highest energy lies about 20 g/cm2 higher in the atmosphere
than expected for a constant elongation rate.
These results are not in contradiction with previous measurements
in this energy range (Wdowczyk 1994; Turver 1992). This elongation
rate, and also the absolute , is
consistent with data at higher energies, obtained mainly by the
Yakutsk and Fly's Eye collaboration (Watson 1997). A constant
elongation rate of 73
g/cm2 from 300 TeV up to 107 TeV (dotted fit
line in the summary diagram 10 in Watson 1997) is an intriguing
hypothesis which is not in contradiction with our data.
6.5. Fluctuation of shower penetration depth
The RMS of the penetration depth distributions - calculated in
reconstructed-energy bins, i.e. biased in favour of the light
component of CR especially at low energies - is shown in Table 3.
It does not show any obvious trend towards a heavy composition.
Therefore the fact that the composition at the highest energy seems to
be heavy with all energy reconstruction methods (Fig. 10) is mainly
determined by the fact that the in
the highest energy bin lies about 20 g/cm2 below a constant
elongation rate.
![[TABLE]](img100.gif)
Table 3. The RMS of the penetration depth distributions [g/cm2] as a function of reconstructed energy (given in the same units as in Table 1) in the data and spectral Monte-Carlo sample. Given is the value inferred for the energy bins as defined with energy-reconstruction method 3, i.e. the specified values contain an A dependent bias. The first error is statistical and the second systematic (due to the systematic error in slope). For the numbers from Monte-Carlo simulations only a statistical error is given. "Mixed composition" represents the expectation for our best-fit chemical composition. Based on numerical experiments the statistical error was taken as the inferred value divided by rather than half of that, as would be correct for Gaussians, due to the non-Gaussian tails of the distribution. The comparison between experimental data and Monte Carlo simulations shows no trend towards a heavy composition at the higher energies.
© European Southern Observatory (ESO) 2000
Online publication: July 7, 2000
helpdesk.link@springer.de  |