Astron. Astrophys. 323, 399-414 (1997)
2. Observations and analysis
2.1. The satellite and the observations
Cygnus X-2 was observed with the Ginga satellite (Makino and
the ASTRO-C team 1987) during 7 observations. For a log of the
observations we refer to Table 1. Due to Earth occultations and
the high background in the South Atlantic Anomaly (SAA) the
observations were broken up into many pieces with a length of several
tens of minutes.
![[TABLE]](img10.gif)
Table 1. Log of the observations
The LAC instrument (Turner et al. 1989) on board Ginga was used in
several different observation modes: the MPC1, MPC2, MPC3 and the PC
mode. In the MPC1 and the MPC2 (hereafter MPC1/2) modes the source was
observed in 48 photon energy channels and the time resolution ranged
from 62.5 ms to 16 s. In the MPC3 mode the source was observed in 12
channels with a best high time resolution of 7.8 ms. In the PC mode
only four overlapping channels were used and the time resolution was 1
or 2 ms. By manipulating the coarse gain (CG) amplifier, which had two
levels, the energy range could be set to either 0.5-37 keV (low CG
level) or to 0.3-18.7 keV (high CG level). By changing the high
voltage (HV) levels the energy range could be extended to even higher
energies (e.g. up to 60 keV). Due to these different CG amplifier
settings, HV level settings and lower discriminator settings (see
Turner et al. 1989) different sets of energy channels were used at
different times.
Generally the satellite attitude control system kept the viewing
direction within 30 arcmin from the position of Cygnus X-2 , which
corresponds to a collimator transmission of
70%. The systematic error in the observed count rates due to
uncertainties in the collimator transmission is of the order of a few
precent (Hertz et al. 1992). Due to, e.g., a long period without
attitude control measurements the offset angle sometimes increased
causing the collimator transmission to decrease to about 40%. The LAC
collimator shows energy-dependent transmission effects due to
reflection of soft X-rays (below 6 keV)
from the collimator walls at large offset angles (Turner et al. 1989).
This low energy reflection was not taken into account when the
intensities were corrected for collimator transmission
(Sect. 2.2), which gives rise to an overestimation of the count
rates for the lowest energy channels. When we encountered data with a
collimator transmission below 60 % we used Figs.
13 and 14 of Turner et al. (1989) in order to correct the count rates
below 6 keV. Usually this means that the count rates were corrected
for an overestimation of 3-4%. Sometimes the offset angle was even
higher and the collimator transmission came below
20%. At such high offset angle the collimator
response is independent of incident photon energy (Turner et al.
1989). However, at these times the fractional uncertainty in the
collimator transmission, and thus in the corrected count rates, due to
aspect jitter is 15% (Hertz et al. 1992). For
this reason, we did not use any data with a collimator transmission
% in our analysis.
2.2. Colour-colour and hardness-intensity diagram analysis
For our studies of the CDs and the HIDs we used data rebinned to a
time resolution of 4 seconds, which were background substracted and
corrected for deadtime and aspect using the ISAS analysis system. The
PC binned data were not corrected for background because no good
background determinations are available for them. The 16 s resolution
MPC1 data of the November 1990 and June 1991 observations had multiple
register overflows and were left outside our analysis. The 4 s
resolution MPC1 data of the June 1987, the October 1989, the May/June
1991 observations and the 2 s resolution MPC2 data of the June/October
1988 and the June 1991 observations had single overflows and the true
count rates could be recovered.
In order to make CDs and HIDs we calculated the soft colour, the
hard colour and the total intensity. The definitions of those
quantities and the energy bands used in order to calculated them, are
given in Table 2. Instead of the colours and intensities, usually
used in the literature, we used the logarithmic value of those
quantities. Any process which influences the intensity in the
specified energy bands by multiplication, i.e. the instrumental
response, produces only a shift in the CD and the HIDs. Using
logarithmic values of the colours and the intensity therefore makes it
easier to compare data of the same source obtained by different
instruments, by simply shifting the diagrams. Also, the
parametrization (Sect. 2.3) in this
way does not depend on the actual values of the quantities or the
count rates, but only on their changes. It can now be used not only
for the CD but for the HIDs as well.
![[TABLE]](img16.gif)
Table 2. The energy bands used to create colour-colour and hardness-intensity diagrams. We used the following definitions of the colours : Soft Colour = 10 log(Band 2 / Band 1), Hard Colour = 10 log(Band 3 / Band 2). The intensity is the count rate in the energy range given in colum 6
Usually the total intensity was taken as the sum of the count rates
in the same energy bands as were used to calculate the colours.
However, for the PC data of the 1987 and the 1988 observations the
total intensity was defined as the count rate in all four energy
channels. For the 1989 PC data the count rates in the three lowest
energy channels were summed, as the fourth channel only contained
background photons. The lowest energy boundary of the 1987 PC data
differs slightly from the lowest boundary in the 1988 PC data
(Table 2). The effect on the Z in the CD and HID is negligible
and therefore we discuss the 1987 and 1988 PC data together. The 1989
PC data is discussed seperately because of the different energy
boundaries (see below).
In order to compare the MPC1/2 data with the MPC3 data we rebinned
the MPC1/2 data to 12 channels using the same energy boundaries as the
MPC3 data. Although in the 1991 observations the MPC3 data had
slightly different energy boundaries compared to the MPC1/2 data
(Table 2), the difference was small and had no significant effect
on the place of the Z in the CD and HID.
Due to the HV and CG levels of 1989 and 1990 observations (energy
range 1-60 keV) the energy boundaries of their spectral channels
differ from those of the other five observations. Due to the broad
energy channels the energy boundaries of these observations could not
be succesfully interpolated to the boundaries of the others.
Therefore, we discuss the 1989 and 1990 observations seperately from
the other observations. In order to get an approximately uniform
analysis we took for the MPC1 mode data colours which resembles the
colours used in the other observations.
2.3. parametrization
Hasinger et al. (1990) introduced the concept of rank number of the
Z track in the CD. This concept was refined by Hertz et al. (1992) and
by Dieters & van der Klis (1996). Here we use the method described
by Dieters & van der Klis (1996) for the measurement of the
position along the Z, with the difference that we perform all
operations on the logarithmic colours and intensity values (see
Sect. 2.2). We therefore transform the two logarithmic colour
coordinates (hard colour and soft colour) into the coordinates
(the distance along the Z track) and
(the distance perpendicular to the Z
track). We also calculated the velocity ( ) and
the acceleration ( ) along the Z track, as
defined as
![[EQUATION]](img20.gif)
where (i) is the position in the Z track on
time T(i), and
![[EQUATION]](img21.gif)
respectively (see also Dieters & van der Klis (1996)). The
length of the NB is scaled to 1 and positions on the HB and FB are
normalised to the length of the NB. The hard vertex corresponds to
= 1, and the soft vertex to
= 2. We applied the
parametrization also on the Z tracks in the HID. We obtained a
different set of and
values which we call
and .
The distribution of or
is a measure for the thickness of the
branches in the CD or HID, respectively. To quantify the thickness of
the branches we calculated the sample standard deviation of the
and
distributions using the method discribed in the Appendix. As a measure
of the overall velocity and acceleration in the CD the standard
deviations of the and
distribution were used, again using the
method outlined in the Appendix.
2.4. Power-spectral analysis
For the power-spectral analysis we used the MPC3 and PC data. We
made FFTs of 128s data segments which resulted in a frequency range of
0.0078-64 Hz (MPC3 7.8 ms resolution data) or 0.0078-256 Hz (PC 1.95
ms resolution data) in the power spectra. We calculated the FFTs for
the energy interval 1.2-18.6 keV (June 1987, June/October 1988 and
May/June 1991 MPC3 observations) or 1.2-19.0 keV (November 1990 MPC3
observation). For the PC mode we used the sum of the energy intervals
0.8-12.3 keV and 0.6-9.1 keV (June 1987), or 0.8-12.3 keV and
0.8-9.1 keV (June and October 1988), or 1.2-17.9 keV and 1.2-15.7 keV
(October 1989).
The average level of the photon counting noise, modified by
deadtime processes (the Poisson level), was estimated and subtracted
using a counter deadtime of 206 µs (MPC3 data) and 16.5
µs (PC data) (see Mitsuda & Dotani 1989). In the
resulting power spectra several components can be identified:
- very low frequency noise (VLFN), which we fitted with
, where is the frequency,
is the power-law index and
the normalization constant.
- low frequency noise (LFN) and a high frequency noise (HFN), are
fitted with
, where is
the power-law index, the cut-off frequency, and
the normalization constant, for the LFN (L) and
HFN (H), respectively.
- HBO, its harmonic, NBO and FBO, are fitted with Lorentzians:
, where the centroid
frequency, the full width at half maximum
(FWHM) of the QPO and the normalization
constant.
The fractional rms amplitudes of the various noise components were
determined by integrating their contribution over the following
frequency ranges ; VLFN: 0.001-1 Hz, LFN: 0.01-100 Hz, HFN: 0.01-100
Hz. We note that variations in the collimator transmission could
influences the VLFN (see e.g. the EXOSAT data of GX 17+2 [Kuulkers et
al. 1996b]). The errors in the parameter values were calculated using
an error scan through space using
. Due to the poor statistics it was not always
possible to fit a HFN component. When HFN was fitted we used
because in Z sources this index was found to be
consistent with zero (Hasinger & van der Klis 1989, Dieters &
van der Klis 1996). The best way to examine the correlations between
the timing behaviour and the position on the Z track is to select a
small part of the Z track, using the
parametrization and determine what the corresponding timing properties
are for that segment. However, our data did not allow this procedure
as we did not have enough data at high time resolution for this
procedure to give sufficient statistics. Therefore we first selected
parts of the data that covered a relatively small
range and afterwards determined an
average value and range for each data
set.
© European Southern Observatory (ESO) 1997
Online publication: June 5, 1998
helpdesk.link@springer.de  |