Astron. Astrophys. 339, 759-772 (1998)
2. Observations and data reduction
2.1. Adaptive optics imaging
G45 was observed in August 1995. ESO's adaptive optics system
ADONIS (Beuzit et al. 1994) was used at the 3.6 m telescope on La
Silla/Chile to obtain high-resolution images in the H and
bands. In , a mosaic of two
frames was made that resulted in a total integration time of 400 s in
the image centre. During the observations, the seeing was
, the high-order adaptive optics correction
improved the full-width half-maximum (FWHM) of the point spread
function (PSF) to 0:004. The integration time for the H image
was 300 s. All frames were subject to standard bad-pixel removal, flat
fielding, and dark-frame subtraction processes before being combined
in the resulting images. For better identification of the point
sources, part of the image was deconvolved with
the point source located at position
( , ) using 200 iterations
of a slightly modified maximum-likelihood deconvolution algorithm. The
modification concerns the last step of each iteration, where the
re-convolution is done with a narrower Gaussian instead of the
original point-spread-function as described in Lucy (1974). This
results in a better resolution of close point sources. Photometric
calibration for the NIR images was obtained by calibrating the fluxes
in the images using separate images of the UKIRT standard Y 4338.
2.2. MIR imaging
Three mid-infrared images were obtained. The first one was taken
using SpectroCam-10 (Hayward et al. 1993) at the 200-inch Hale
Telescope of the Palomar Observatory
2. The filter
effective wavelength was 11.7 µm with
µm. A 5 frame mosaic was combined
into the image, the average on-source time at each pixel is 20 s. The
star Lyr served as a standard for flux
calibration. This image will be referred to as the 12 µm
image in the discussion. The second image was taken using MANIAC
(Böker et al. 1997) at ESO's 2.2 m telescope on La Silla/Chile.
MANIAC's N-band filter was used with
µm and
µm. The total integration time
sums up to 1135 s for the frames that were combined into the image.
Photometric calibration was obtained by observing the standard stars
Aqu, Ser, and
Gru. This will be our 10 µm or
N image.
A third MIR image was obtained in the L-band
( µm,
µm). The new MIR-Camera TC-MIRC
(Robberto et al. 1994) was used at the 1.5 m TIRGO telescope on
Gornergrat/Switzerland. The total integration time for this image was
600 s. Photometric calibration was achieved by observing the standard
star HD203856. This image will be referred to as 3.8 µm
or L image.
2.3. Narrow band imaging
The Br image was taken using BLUE-MAGIC
(Herbst et al. 1993) at the 2.2 m telescope on Calar Alto/Spain which
belongs to the German-Spanish Astronomical Centre. One image was taken
using the Br
( µm,
µm), another one with the adjacent
continuum filter ( µm,
= 0.022 µm). The total integration
time for each of these images was 900 s. Photometric calibration was
achieved by using the standard star GL748. Its magnitudes at the
wavelengths of the continuum and the Br filter
were derived by interpolating its known spectral energy distribution,
assuming a featureless spectrum. After subtraction of the continuum,
it turned out that the residual total fluxes in the stellar PSFs in
the image were below the sky sigma level, so no further corrections
were applied.
2.4. The astrometric reference frame
When drawing conclusions on the nature of a certain object, a great
deal of the information is taken from the comparison of its structural
appearances at different wavelengths. Apart from other issues like the
different beam sizes for each observation, it is crucial to establish
a common astrometric reference frame for all images. This is
especially difficult when one tries to match images of 0:002
resolution like our deconvolved image and the
6 cm VLA maps from WC89. We adopted a two-way strategy to build our
reference frame: First, the position of the wavefront calibrator which
is visible in the image at position
( , ) (see Fig. 1) was taken
from the digitized sky survey. The image was then calibrated using the
known scale of 50 milliseconds of arc per pixel. This tied the NIR
reference frame to the digitized sky survey. Secondly, the astrometry
for the Br image taken with MAGIC was obtained by
optimizing the cross-correlation with the 6 cm VLA image, the latter
one being smoothed to the same resolution before the procedure. As the
Br image contains the same stars as the
image (before continuum subtraction, of course),
we were able to check the two astrometric frames against each other.
It turned out that the deviations between them were below 0:002. This
indicates that all larger offsets are indeed real. The method of
optimizing cross-correlation was also applied to tie the
10µm image to the reference frame. Before optimizing the
cross-correlation between the images, a mask was applied that blanked
everything except the ionization front. The reference frames of the 12
and 3.5 µm images were chosen to make the MIR
point-source at position ( ,+10:005) appear at
identical positions in all MIR images.
2.5. Photometry
Photometry of the point sources detected in the AO images was
performed in an aperture of 0:005 diameter because of the crowdedness
of the field. Due to the non-Gaussian shape of the AO-corrected PSF,
an aperture correction of 0.95 mag in and
1.23 mag in H was applied to account for flux outside the
aperture. This correction was determined using sources at the
reference position as well as a and b (in H,
a is not visible). Background subtraction was achieved by
subtracting the mean value of six sky measurements around the
corresponding source using the same aperture to account for the
varying background. The detection limit in the
-image is mag, in the
H-image it is mag, both derived from the
sky noise. Photometric accuracy is generally better than
mag. Photometry of the MIR images was done
using an aperture of (2:0075 at
3.5 µm) diameter. Background subtraction was achieved in
the same way as in and H. The accuracy of
the photometric measurements in these images amounts to approximately
7% of the flux level or 0.1 magnitudes at most. The total source flux
was determined in these images as well as in the
image by applying an aperture of diameter.
2.6. A word on image filtering
All images presented as results from our imaging campaign were
subject to filtering with the multi-scale maximum entropy method
described by Pantin & Starck (1995). This method uses a wavelet
decomposition to detect and remove the noise from an image. The
band image was also subject to this kind of
filtering before the deconvolution was applied. However, all this
filtering was done solely for illustrative purposes. All photometric
and mathematical operations described throughout the paper were
performed on the "raw" data, before filtering or deconvolution. It
should be stressed that all features discussed in the text can be
found in the raw data. Therefore, we are certain not to describe any
filtering artifacts whatsoever!
© European Southern Observatory (ESO) 1998
Online publication: October 22, 1998
helpdesk.link@springer.de  |